Abstract
Purpose of the study: To comparatively evaluate the effect of height differences on the retention and wear behaviour of ball attachment system over mandibular dual implants.
Materials and methods: Two groups of twenty samples each containing polymerised master block of dual implant analogues at same height in Group 1 and dual implant analogues of different heights in Group 2 were fabricated along with ball abutments, nylon cap and metal housing. Qualitative analysis of samples for wear factor were carried out using stereomicroscope and quantitative analysis of retention force at baseline, 360, 720, 1080, 1440 insertion-removal cycles were determined using Universal testing machine. statistical analysis were carried out using independent ‘t’ test and paired sample ‘t’ test.
Results: Group I test samples, the sample mean obtained at baseline, 360, 720,1080, and 1440 cycles were 24.73N, 23.49N, 22.46N, 21.07N, & 19.28N respectively. Group II test samples, the sample mean obtained at baseline, 360, 720,1080 & 1440 cycles were 27.13N 25.80N, 24.16N, 22.73N and 21.85N respectively. comparison of the mean retention force of the test samples at baseline, 360 (3 months), 720 (6 months), 1080 (9 months) and after 1440 (12months) insertion-removal cycles within Group 1 and Group II using Paired sample ‘t’-test was statistically significant at all levels.
Conclusion: Retention force was found to be higher on dual implants placed at different heights in mandibular anterior region and is statistically significant and corroborated by the Lesser surface deformations of attachment system.
Keywords: Dental implants; Denture retention; Mandibular prosthesis; Prosthesis retention; Titanium
Introduction
Edentulism is considered as a poor health outcome and it often compromises the quality of life of edentulous people [1]. Complete edentulism is a condition that is most prevalent among the elderly population [1,2]. It has a negative impact on the masticatory function, speech, esthetics and on the quality of life. The prosthetic management of the completely edentulous patient has been a major challenge in dentistry [3,4]. Conventional complete denture have been a standard and classical treatment option for the completely edentulous patients for more than a century [1,3,5]. Conventional complete denture is commonly provided to completely edentulous patients to restore masticatory function and esthetics [1,5]. The complete denture wearers often encounter problems like difficulty in mastication, discomfort during speech along with poor stability and retention [1,5,6]. They are usually satisfied with the upper denture but the majority of them often struggle with the lower denture due to the lack of retention and stability [1,7].
Alveolar ridge resorption is one of the important factors which is associated with loss of stability and retention in lower denture. It also reduces the amount of underlying tissue available for denture support [1,7]. The resorption of the residual alveolar ridges is a chronic, continuous, life-long catabolic process of bone remodelling. The rate of reduction in size of the residual ridge is maximum in the first three months and then gradually tapers off [8]. According to Boucher, during the first year after tooth extraction, the reduction in the residual ridge height in the mid sagittal plane is 2-3mm for maxilla and 4-5mm for mandible and the annual rate of reduction in ridge height is 0.1-0.2mm for mandible and this reduction in height of the ridge is four times less in the maxilla [1,8,9]. Many studies have shown that the poor retention and stability could be managed using a fixed prosthesis supported by five or six implants or by the fabrication of an overdenture to implants [9]. However, placing six implants in an atrophic mandible is not always possible. Therefore, the concept of using two or four implants to support an overdenture was introduced.
Overdenture is a therapeutic approach which is directed in improving the oral function in elderly edentulous patients [10]. The concept of overdenture initially involved fixing mechanical attachments to teeth roots to enhance retention and stability of conventional complete dentures [6,9]. With the evolution of implant overdentures, there is reduction in displacement of the prosthesis due to lateral forces leading to better retention and stability improving the masticatory function and overall quality of life [5,11,12]. Crum, Rooney [8] reported an average of 5.2mm loss in the alveolar ridge height in denture patients compared to overdenture patients where there is 0.6mm loss which is relatively low. According to McGill consensus, two-implant overdentures has been accepted as the gold standard treatment care for the edentulous mandible [5,13]. Various factors contribute to the success of an implant supported overdentures, including the fit and precision of dentures and the retentive capacity of its attachment system to provide long duration of function. Therefore, Retention is considered as one of the most significant factors in determining patient satisfaction in removable prosthodontics [9,14] and is defined as the force that resists withdrawal along the path of insertion and stabilizes the overdenture during its function [14,15].
Also, the choice of attachment is primarily dependent on the retention required, jaw morphology and anatomy, condition of the mucosal ridge, oral function and patient compliance for recall appointments. There are different attachments to retain implant supported overdentures. They were classified on the basis of their flexibility, geometrical shape and cross section, casting precision and manufacturing procedures. In one of the major classifications, these attachments are divided into splinted and unsplinted systems. Splinted systems includes the bar attachments whereas unsplinted systems includes magnets, ball types and locators [5,7,16,17].
Within these systems, self-aligning attachments (locators) and ball attachments have been frequently used due to their simplicity. Specifically, ball attachments are considered as one of the simplest type of attachments for clinical usage. It provides varying degrees of resiliency in both vertical and horizontal directions. Also, the specific design of the ball attachments may influence the amount of its free movement thereby limiting its resiliency [1,5,11,18]. The retention and the longevity of the attachments have been studied as a common problem in many clinical and in vitro studies. There are lot of studies evaluating the retentive force of different attachment systems in mandibular two-implant overdentures simulating different periods of clinical usage [6,7,11,19]. Some studies have comparatively evaluated the retention capacity of ball, bar and locator attachments in implant overdentures [19]. Studies have also compared retention between locator, ball and magnetic attachments [20].
Many studies have compared the retentive capacity of different color codings of a single attachment system [14,21]. Few authors have studied the retentive forces of the attachments by placing implants at different angulations [14,20]. Previous studies have evaluated the retentive force using micro material testing machine (MTM) [9,15], a dental mastication simulator [14,21], CS-Dental testing machine [16], load cell (cyclic fatigue machine) [22], Imada device (IM) [23,24] and Universal testing machine (UTM) [6,11,24] and among these methods Universal testing machine (UTM) has been accepted as reliable and valuable instrument to test the retention forces in vitro [6,11]. One of the main factors of retention loss is the change induced on the components of the attachment systems as a result of wear [6,11]. The factors which are associated with the clinical wear of attachments includes masticatory forces, parafunctions, temperature and composition of saliva, products used for the maintenance of denture and the presence of food residues [22]. The wear of overdenture retentive mechanisms has been identified as the most common prosthodontic complications which is about 33% [2].
In addition to the retentive forces, few studies have also evaluated the amount of wear and its effects in the attachment systems. Wear have been assessed using Metallographic microscope [25], Micro material testing machine (MTM) [15,26-28], Coordinate measuring machine (CMM) [18], Scanning electron microscopy (SEM) [14,29], and stereo microscopy [6,17]. The positioning of the implant overdenture attachments is very important for two-implant overdentures. This is because during pathological overloading, the bone around the implants becomes deformed and resorbed due to the increased stress and strain gradients [5]. This situation may cause incompatibility of the components of the implant system and microfracture of the implant [5]. Clinicians predicted that the two independent implants must be positioned at the same occlusal height, parallel to the occlusal plane [24]. However, placement of implants in the same occlusal height in completely edentulous patient is not always possible since the alveolar bone resorbs with different types of resorption patterns [5]. The rate of residual ridge resorption differs from person to person and even at different times and sites in the same person. Unlike in maxilla, the speed of bone loss in mandible is different in different parts of the jaw, distal parts of the residual ridge disappear faster than the anterior parts, which may affect the symmetry of the remaining bone height resulting in difficulty in placing the implants in position [8].
Few studies have evaluated the retention force of attachment systems retained by single or two implants and also at different abutment heights [30]. Ozan O and Ramoglu S have compared two different attachment systems in two-implant overdentures by evaluating the stress distribution in peri-implant site and on attachments by positioning implants in different height levels using the 3D Finite Element Analysis method [5]. Currently, studies comparatively evaluating the retention and wear behaviour of two-implant overdenture attachments with implants positioned at different height levels are lacking. In light of the above, the aim of the present in vitro study was to evaluate the effect of implant height differences on the retention and wear behaviour of ball attachment system in mandibular two-implant overdentures. The null hypothesis for the present study was that implant height difference does not affect the retention and wear behavior of the dual implant overdenture attachments.
Materials and Methods
A rectangular block of dimensions 60mm x 20mm x 10mm, made of plaster of paris (Ramaraju Mills ltd., India) was custom made to serve as an index for the fabrication of wax blocks of similar dimensions and then to be converted into heat cure acrylic resin blocks of uniform dimensions to be used in this study. Addition silicone impression material of putty and light body consistencies (Aquasil, Denstply, USA) were used for obtaining the index in a single step procedure. The putty was hand mixed with equal quantities of base and catalyst to obtain homogenous dough. Light body material in a cartridge was attached to the auto mixing gun (Heraeus Kulzer, Dormagen, Germany). A spiral mixing tip (yellow 70mm , Adenta, USA) was attached to the cartridge tip and material was injected gently over the custom- made plaster block. The mixed putty was also placed over the plaster block and left undisturbed until set. After setting, the plaster block was removed from the index and the mold space area was inspected for defects and acceptability. The putty index thus obtained was used to fabricate the test samples of standardized dimensions for this study.
a) Master wax blocks(n=2),
b) Prosthetic wax blocks (n=20).
Modeling wax (Hindustan manufacturer, Hyderabad, India) (was melted and poured into the mold space created by the putty and was allowed to cool. After the wax had completely hardened, the wax blocks were retrieved carefully and placed at room temperature. Twenty-two such wax blocks of 60mm x 20mm x 10mm dimensions were fabricated in which two blocks were used as master blocks and twenty blocks were used as prosthetic blocks to conduct the study. Out of the two master blocks, in one block a height of 2mm was increased with wax on one half of the block exactly measured from the centre of the block. Similarly, out of the twenty prosthetic blocks, ten blocks were increased to a height of 2mm as it was done for the master blocks. Two implant analogs with the diameter of 4mm and height of 12.7mm (Norris dental implants, Israel) were positioned parallel to the insertion-removal path and to one another, on each of the master blocks using a dental surveyor (seshin precison Ind., co, korea). The master blocks were placed on the surveying platform of the surveyor and stabilized. The surveying platform was made parallel to the floor. The implant analogs were positioned into the wax blocks at a distance of 22mm from each other with equal distance from the centre of the block. The implant analogs were submerged into the wax block up to the crest module of the implant analogs (2 Master blocks with implant analogs and 20 Prosthetic blocks).
The two master blocks with the implant analogs and the twenty prosthetic blocks were fabricated using heat polymerized acrylic resin. The ball abutments of diameter 2.5mm and height 2mm (Noris Dental implants, Israel) were screwed into all the implant analogs embedded in the master blocks. Modelling wax was used to block the undercuts around the ball abutments in the master blocks. The metal housing of 5mm diameter and 3.2mm height along with nylon cap insert of standard retention (Noris Dental Implants) was placed over the ball abutments for picking it up onto the prosthetic block. The prosthetic blocks were now drilled with acrylic burs to create space for the metal housing with nylon cap inserts (Figure 1). In order to prevent the acrylic adhesion of two blocks, a thin layer of separating medium (Tejpal pharma and surgical, India), is applied on the master blocks. Clear autopolymerising acrylic resin (RR cold cure, DPI, India), was poured into the space created and the metal housings with the nylon cap insert placed over the ball abutments in the master blocks were picked up in the prosthetic blocks. Upon setting, the master block with ball abutment was retrieved from the prosthetic blocks with the metal housing and nylon cap insert. Twenty prosthetic blocks thus obtained were divided into two groups of ten blocks each according to the height difference.
1. Group I (n =10) Test samples for the implant analogs placed at same height.
2. Group II (n = 10) Test samples for the implant analogs placed at different height.
Each test sample was placed over the imaging platform of the stereo microscope prior to retention testing. Images were captured in a computer-controlled software system in 40x magnification at an object lens distance of about 13mm and the same magnification had been used for all the test samples (Figure 2). A total of twenty test samples were tested individually in the Universal testing machine (INSTRON 8874) (Figure 3) to measure the force required to separate the prosthetic block from the master block. Both the master block with ball abutment-implant analog assembly and the prosthetic test samples were positioned on the machine table and secured tightly into the upper and lower clamps of the universal testing machine. Engagement and disengagement of the attachments were carried out at right angles to the horizontal level of the blocks. The testing machine was programmed to apply 1440 cycles of insertion-removals. Assuming that a patient removes and inserts his prosthesis four times a day, the retention force values were noted at baseline, after 360 cycles (simulating 3 months of clinical use), 720 cycles (simulating 6 months of clinical use), 1080 cycles (simulating 9 months of clinical use) and after 1440 cycles (simulating one year of clinical use) The test samples were kept moist with artificial saliva throughout the testing as it acts as lubricant to simulate potential in-vivo conditions. The tests were conducted in an open room at room temperature.